Angel of Water

 


Angel of Water

Angel of Water® US Distributors- NIA, Inc. | 174 Summit Ave., Suite 203, Summit NJ 07901 | (800) 834-9945 or (908) 451-5748 | Fax (908) 598-7449 | jim@ctnf.org

 

 

Control Switch Panels
The Angel of Water® Surround has a tower switch panel with LIGHTS, UV, CYCLE, and FLOW switches and
a remote switch panel with a FLOW switch and an INDICATOR LIGHT.
The FLOW switches on both panels must be in the ON position for the function to operate. If one function
switch is turned OFF, then that function will not work.
The switches are described below:
LIGHTS - (tower control panel only) - Turns on all lights for the inside and outside viewing tube areas and
tower cabinet to illuminate tank water level and make it visible through the viewing window. This switch also
activates all other switches. When this switch is ON, all other switch functions will work. When this switch is
OFF, none of the other functions will work.
UV - activates ultraviolet light in UV Filter. This should only be turned ON just prior to water flowing to
the person on the System. Turn OFF when session stops for more than a couple of minutes or when session
is finished. Turn ON again as session resumes. The UV light should NOT be left on continuously.
CYCLE - activates the water cycling device to move water from main tank to recirculating tank. This keeps
water temperature evenly mixed throughout the use of the entire contents of the water tank and maintains
an even and constant stream of gravity-fed water to the patient. The water cycling device should only be ON
when water is flowing through it. When water in large main tank runs out, turn CYCLE switch OFF until tank
is filled with more water. Then resume session with CYCLE switch ON to maintain even and constant stream
of gravity-fed water.
FLOW - activates Solenoid Valve that allows water flow from tank into the patient. A small round indicator
light next to the FLOW switch on the remote control panel lights up and is visible when water is flowing to
patient.
The Temperature Controller and Display with Sensor is described below:
The Temperature Display on the Angel of Water® CM-1 Surround shows the water temperature and is located
on the front of the tower cabinet, visible to both the patient and healthcare practitioner.
The temperature controller with sensor senses water temperature and automatically closes Solenoid Valve during
Regular Mode of Operation should the temperature of the water exceed 103 degrees Fahrenheit (39.8 C.).
Should the Solenoid Valve shut off due to water temperature exceeding the acceptable range (99 – 103 degrees
F. and approximately 37-39 degrees C.), the healthcare practitioner should simply add cool water to the
main tank by turning the Mixing Valve on the basin console (within reach on right topside of basin) to a cooler
setting and then adding water to the tank by opening Valve #1. Water may have to be let out first by opening
Valve #2 to accommodate more water. Once the temperature returns to the safe range (102 degrees F. or less)
(38.8 C.), the Solenoid Valve will automatically reactivate, allowing the FLOW function to deliver properly- tempered
water to the patient.

 

 

 

WATCH DEMO

Medical Opinions | BUY NOW | Installation | Starting a Colon Therapy Business | Pictures | Accessories/Supplies | User Manual | Training/ Certification

Conducting a Session | Routine Maintenance | Cleaning Unit | Trouble Shooting | Features | Control Panel Switches | Valve Settings

SAMPLE FLOOR PLAN | DESIGNING YOUR ROOM | BLANK FLOOR PLAN | ARIAL VIEW

 

Angel of Water® US Distributors- NIA, Inc. | 174 Summit Ave., Suite 203, Summit NJ 07901 | (800) 834-9945 or (908) 451-5748 | Fax (908) 598-7449 | jim@ctnf.org

BACK TO HOME PAGE